Research note / Cycle 024

Reading the parts where they stand

· Nadia Okafor, Tomás Ibarra

A bench meter with no calibration record reads AQ-RC01’s capacitor and resistor in the board. They sit near their labels, a lifted leg changes nothing the display can show, and the January records could not have told the difference.

Two charts of the January display readings with curves calculated from the February 26 component readings. The left chart shows the rises and the fall; the right shows the rise with the 1 megohm shunt levelling near 0.90 volts. The points lie on or close to the curves.
January 23 and 26 display transcriptions and curves built from the February 26 AQ-M02 readings, not fitted. The curve from the labels differs by at most 0.004 V and is not drawn.

An instrument the decision assumed

Aletheia’s amplifier will need an acquisition chain whose parts are known before any signal is interpreted. Its test object is still AQ-RC01, a resistor marked 100 kΩ and a capacitor marked 47 µF on a breadboard, read with the handheld meter AQ-M01. On February 21, in Two rises we already had, the program decided to measure the resistor, the capacitor and the meter’s input before comparing any more curves. The markings had been read. Nothing had been measured.

That decision assumed an instrument. On February 25 Okafor and Ibarra checked what the workshop holds. AQ-M01’s front panel and the workshop’s copy of its manual list voltage functions only, so it cannot read either part. The workshop register lists a mains-powered bench multimeter, WS-DMM-2, with resistance and capacitance functions, and it was checked out to the program that morning as AQ-M02. The calibration notebook has one entry for it: a function check on September 14, 2026, recording that all ranges respond, with no reference values or uncertainty. No calibration certificate is on file.

What the displays can be said to show

The manual gives accuracy figures that apply within one year of calibration. Without a calibration record they are not ours to use. Every value below is what AQ-M02 displayed, at its resolution, and none carries an accuracy or an uncertainty.

Keysight’s guide to digital multimeters describes how such a meter reads capacitance: it charges the part with a known current and computes the value from the change in voltage. Keysight’s note on capacitance meters adds that a multimeter measures under a constant DC signal and cannot give equivalent series resistance, and that dedicated instruments test at chosen frequencies such as 120 Hz and 1 kHz. The capacitor’s ±20% tolerance, discussed in Two rises we already had, is defined at 120 Hz. A DC reading is a different measurement of the same part, not a check of that tolerance.

Taking a part out, or not

Fluke’s guide to measuring capacitance with a multimeter says to switch off power, discharge the capacitor and detach it from the circuit to avoid measurement errors. Detaching it would end the arrangement that produced every January reading. Before touching anything, Ibarra drew the board as it stood and Okafor checked the drawing against it. A toggle switch connects one end of the resistor either to the source’s 1 V lead or to the return rail. The resistor’s other end meets the capacitor’s positive leg, and the negative leg sits in the return rail. AQ-M01 clips across the capacitor during runs.

The drawing showed a shorter route. With the source lead unplugged at the source and the meter unclipped, the switch’s 1 V position leads nowhere, and nothing else connects across the capacitor. On February 26 the source was switched off and the capacitor discharged for sixty seconds through the resistor. AQ-M01 then read 0.00 V across it before its leads were removed. With the switch at 1 V and AQ-M02 on resistance, the display across the capacitor rose without settling and showed OL on the 20 MΩ range after about fifty seconds. Nothing below that range bridges the part. Leakage above it is not excluded.

At 20 °C on the workshop’s uncalibrated wall thermometer, the capacitor read 48.4, 48.2 and 48.2 µF in the board, with the switch returned to the return rail for thirty seconds between readings.

One leg, lifted and put back

Ibarra asked for one more check: the same part with nothing else connected, which is what the published method assumes. Okafor approved it once the in-board readings were written down, on three conditions. Lift one leg only, return it to the marked hole, and record the board afterward as reassembled. At 11:12 Ibarra lifted the negative leg out of the return rail, and the part read 48.2 and 48.2 µF. At 11:24 the leg went back beside the pen mark, and the board read 48.3 and 48.2 µF. For this quantity the lift changed nothing the display could show.

It ended something else. The arrangement used on January 23 and 26 and February 3 stopped at 11:12. The same parts are in the same holes, but the reinserted leg makes a new contact, and from 11:24 the board is recorded as AQ-RC01 reassembled February 26. Any later run on it is a new session, not a repeat of January.

The resistor and the meter’s input

The series resistor read 98.62 kΩ in the board. With the source lead open, nothing parallels it. AQ-M01’s input was not read directly, because an ohmmeter’s reading of another meter’s input depends on that meter’s own circuitry.

A cheaper route was already on file. On January 26, in How much circuit does the meter add?, a 1 MΩ and a 10 MΩ resistor were each placed in series with the meter, which read 0.91 V and 0.50 V. Neither resistor was measured then. Both were still in the labelled bag from that session, together with the 1 MΩ shunt used on January 23. Clipped in the lead jaws, they read 0.9937, 10.210 and 0.9981 MΩ. A first attempt on the 10 MΩ part, holding the leads by hand, read low and drifted, and it was discarded. With the two series values, the January 26 readings put AQ-M01’s input between 9.91 and 10.52 MΩ, allowing only for the display’s rounding of each reading and of the 1.00 V source.

The January runs, set against the parts

Analog Devices’ transient-response activity gives the time constant as the capacitance times the resistance seen from the capacitor’s terminals. With the capacitor at 48.2 µF, its most common display, the resistor at 98.62 kΩ and the meter at 10.21 MΩ, that gives 4.708 s, settling at 0.990 V. Taking the other capacitor displays and the whole input range moves it by about 0.02 s. Nothing is fitted. Every value comes from a display or a January record.

The resulting curve lies within 0.010 V of every January 23 rise reading, within 0.006 V of the January 26 rise and within 0.009 V of the January 23 fall. A fourth run had never been compared with anything: the January 23 rise with the 1 MΩ shunt across the capacitor, which levelled at 0.90 V. With the shunt’s read value, the same parts give 4.288 s, settling at 0.902 V, within 0.009 V of all six readings. A capacitor 20% below its label, the edge of its tolerance, would miss the 5 s readings by more than 0.05 V.

What the readings do not settle

The curve from the labels, which Analog Devices’ introductory activity works as its own example of 4.7 s, never differs from the curve from the read values by more than 0.004 V. That is less than one step of AQ-M01’s display, so the January records could not have told the labels from the parts. Two rises we already had asked whether the closeness to the nominal curve was a property of these parts or a coincidence. As far as one uncalibrated meter can say, it is a property of the parts. The capacitor reads 2.6% above its label and the resistor 1.4% below. Every January run depends on the two mainly through their product, so errors of equal size and opposite sign would agree as well.

The early readings still sit slightly below every curve: 0.18 V against 0.190 V at 1 s on January 23, and 0.35 V against 0.356 V at 2.1 s on January 26. A start a tenth of a second late would do that, and so would reading early. Only a recorded step edge can separate them.

Decision

The displays are recorded as AQ-M02 readings without calibration, and they replace the labels in later comparisons with that status stated. Before any of these numbers is quoted with an uncertainty, the program needs a second instrument or a reference with a documented value. The two-channel capture specification follows. AQ-RC01 reassembled February 26 is the board for any new run.

No time constant is fitted and no calibration is claimed. No amplifier, participant or implant is involved, and nothing here bears on the program’s ET/ED questions.

Results at a glance

QuantityMarking or January assumptionRead February 26 (AQ-M02, no calibration)Status in the January comparison
Capacitor47 µF, 25 V; no tolerance mark48.2 µF; seven displays 48.2–48.4 in the board, lifted and reinsertedRead value
Series resistor100 kΩ, 5%98.62 kΩ in the boardRead value
AQ-M01 inputNot marked; 10 MΩ modeled since January 23Not readInferred, 9.91–10.52 MΩ, from the January 26 dividers
Divider resistors1 MΩ and 10 MΩ bands0.9937 and 10.210 MΩ, looseRead values, used for the inference
Shunt AQ-L011 MΩ bands0.9981 MΩ, looseRead value, loaded rise only
Time constant, unloaded runs4.65 s from labels and modeled inputNot readCalculated 4.708 s; never observed directly
January displaysTranscribed January 23 and 26UnchangedCompared: largest difference 0.010 V